Acidifying GP for digital photometers – 10 sachets
Acidifying GP by Water-i.d. for sample preparation for high-range chlorine testing with a compatible digital photometer. The sales unit contains 10 sachets. Use with the other reagents specified by the method; this reagent alone is not a complete test. See the product description for PoolLab 2.0 instructions and required equipment. For water analysis only.
Acidifying GP: reagent for sample preparation for high-range chlorine analysis
Acidifying GP by Water-i.d. is a sample-conditioning powder used before high-range chlorine testing. Replace the specific reagent without buying a new testing instrument. Reliable results require the correct program, sample preparation and reagent combination.
What is included
10 sachets of Acidifying GP. The price and order quantity refer to this sales unit, not to the manufacturer’s wholesale box.
Other required reagents / equipment: Chlorine HR (KI). Other reagents and the photometer are not included.
| Parameter | Value |
|---|---|
| Manufacturer | Water-i.d. |
| Form | Powder in individual sachets |
| Sales unit | 10 sachets |
| PoolLab 2.0 method | CLHR |
| PoolLab 2.0 range | 5–200 mg/l Cl₂ |
| Other reagents / equipment | Chlorine HR (KI) |
Quick measurement instructions – PoolLab 2.0
- Measure 10 ml of sample water into a clean chamber. Close the cover and perform ZERO without reagents.
- Select Chlorine HR (KI), labelled CLHR. Add the entire contents of 1 Acidifying GP sachet and dissolve completely.
- Add the entire contents of 1 Chlorine HR (KI) sachet and stir until completely dissolved.
- Close the cover, start OK/TEST and wait until the countdown finishes. Read the result in the units of the selected method.
Do not skip the instrument countdown. Use clean equipment and rinse the chamber and stirring rod thoroughly after testing. Check exact conditions, pretreatment and possible interferences in the manufacturer’s instructions.
Digital photometer compatibility
The stated procedure and measuring range apply to the corresponding PoolLab 2.0 method. For another photometer, use this reagent only if its manufacturer specifies this exact reagent and procedure. A matching parameter name alone does not establish compatibility. Do not confuse No. 1 with No. 2 or LR (low range) with HR (high range).
Storage and handling
Keep in the original tightly closed packaging as directed by the manufacturer. Prevent contamination, check the expiry date and follow the label and safety data sheet for the supplied batch. For analysis of a water sample only; do not add directly to pool water. Keep out of reach of children.